Clarus SQ 8 MS Hardware Guide
128
Vacuum System Maintenance
Vacuum system maintenance consists of the following:
•
Checking the forepump to ensure the oil is at the proper level.
•
Adding oil to the forepump reservoir.
•
Replacing forepump oil.
•
Replacing foreline trap pellets.
Complete pump instructions are in the instruction manual supplied with the pump.
Maintanenace of the Turbomolecular Pump
You should never service the turbomolecular pump. Call your PerkinElmer Service
Representative for the maintenance and any problems you may have with the pump.
Checking the Forepump Oil Level
1.
Locate the oil level indicator window on the forepump.
0
l
ll
0
Mass Spectrometer
Connection Port
Exhaust Port
Oil Filler Plug
Drain Plug
Oil Level
Indicator
Gas Ballast
Switch
On/Off
Switch
Mode Selection
Switch
Max
Min
Voltage Selection Switch
(on the side)
Figure 42 Location of the oil level indicator.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......