Clarus SQ 8 MS Hardware Guide
126
3.
Replace with a new electron multiplier by sliding it forward toward the source
and clipping it back into place.
4.
Return the ion optics assembly. See Removing and Returning the Ion Optics
Assembly on page 114.
Cleaning the Prequads
CAUTION
ATTENTION
To prevent contamination of the mass spectrometer, always wear clean,
lint-free, powder-free nitrile
,
nylon,
or PVC gloves before touching,
removing or replacing parts. Never touch these parts with ungloved
(bare) fingers, as this will introduce contaminants into the system.
Pour éviter toute contamination du spectromètre de masse, porter
toujours des gants de nitrile, de nylon ou de PVC sans pelu, sans
peluches, sans poudre avant de toucher, d'enlever ou de remplacer les
pièces. Ne touchez jamais ces pièces avec des doigts non agglomérés,
car cela introduira des contaminants dans le système.
1.
When operating under normal circumstances, you may not have to remove the
prequads from the ion optics assembly but you will need to access the optics
assembly. See Removing and Returning the Ion Optics Assembly on page 114.
Figure 41 Location of the prequads in the optical assembly.
Prequads
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......