Problem
Probable Cause
Solution
The ultimate pressure is
poor.
Is the cooling inadequate?
Check the cooling-air flow and
correct if possible.
Check the cooling-air duct for
obstructions and correct as
necessary.
If the cooling air flow is fine
and there are no obstructions
contact your PerkinElmer
service representatives.
Is the backing pressure
high?
Check for a leak in the backing
pipeline and poor backing
pump performance. Correct as
necessary.
The pump is very noisy or
there is excessive vibration
or both.
Is the noise irregular and
getting progressively
worse?
If so, a bearing may be
defective. Contact your
PerkinElmer service
representative.
Is the pump making a
constant high pitched
noise?
If so, the rotor may be out of
balance. Contact your
PerkinElmer service
representative.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......