Clarus SQ 8 MS Hardware Guide
124
8.
Using any Allen key, as shown in the following figure, pull up on the underside
of the Sliding Lock attachment to lock the connector into place. See the
following figure.
Figure 39 Sliding Lock attachment shown in the “up” (or locked) position.
9.
Replace and tighten the ferrule and nut for the reference gas vacuum hose.
10.
Secure the optics assembly in place with the four hex screws previously
removed.
NOTE: Do not tighten these hex head screws beyond finger-tight. The vacuum will pull the
ion optics assembly top plate onto the large o-ring to create the seal. Over-tightening
these screws may cause the ion optics assembly top plate to warp, which may cause a
leak.
11.
Secure the top cover by securing the slots in the cover to the tabs on the
instrument.
12.
Use the medium sized Philips screwdriver to reattach the two screws on to the
left access panel.
13.
Replace the source. See
Removing and Returning the Source
on page 82
Bookmark not defined.
.
Allen Key
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......