Troubleshooting
163
Problem
Probable Cause
Solution
Dirty injector liner.
Clean or replace.
Column has active sites.
Equilibrate or replace.
Chromatographic peaks
too wide.
Injector too cool.
Raise injector temperature.
Sample overloading the
column.
Use a split injection or a smaller
sample.
Incorrect GC oven program.
Enter a new oven program.
Discrimination of relative
peak intensities.
Poor resolution or improper
tuning.
Retune.
Unstable filament.
Replace filament.
Poor calibration.
Recalibrate.
Air leaks at detector.
Check He/Air ratio.
Peaks are flat-topped.
Signal strength exceeds
dynamic range of detector.
Reduce multiplier voltage.
Sample is too strong.
Dilute or split.
High baseline.
Dirty sample.
Prepare and filter a new sample.
Air leak at injector.
Locate and fix the air leak.
Contaminated carrier gas.
Replace the gas tank.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......