Clarus SQ 8 MS Hardware Guide
158
Problem
Probable Cause
Solution
Unusually high emission
setting.
Ion volume dirty.
Clean the inner source.
Prefilters dirty.
Clean the prefilters.
Poor tuning.
Retune.
Tuning peaks show
precursors (forward slope
shoulders).
Poor tuning.
Retune.
Dirty prefilter.
Clean prefilters.
Dirty source.
Clean source.
Particulates on analyzer
rods.
Clean the particulates of the rods
with a methanol-dampened lab
wipe.
Blow the particulates off the rods
with helium or dry nitrogen.
Defective or damaged
analyzer.
Call a PerkinElmer service
engineer.
No reference peak.
Reference gas off.
Turn on the reference gas.
Empty reference vial.
Visually check and refill.
Faulty solenoid.
Listen for click when
activating/deactivating the valve.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......