Clarus SQ 8 MS Hardware Guide
106
6.
The defective filament assembly will fall out of place, so carefully hold or
catch it with your other hand.
Figure 25 Removing the defective filament assembly.
7.
Position the new filament assembly with the filament side up.
8.
Position the new filament assembly with the coiled filament wire facing and
centered in the orifice in the ion volume and the white ceramic rests on the tab.
9.
Insert the Filament retaining clip into the slot in the block to hold it in place.
10.
Align the four male gold leads (Trap, Repeller and Filament) with the four gold
Filament