Troubleshooting
153
Problem
Probable Cause
Solution
Turbo pump will not
accelerate
Pump malfunction.
Call a PerkinElmer service
engineer.
Pump controller
malfunction.
Call a PerkinElmer service
engineer.
GC is not properly
configured.
Set the proper GC
configuration for your site.
Vacuum light continues to
blink.
Large leak.
Locate vacuum leak and
correct.
Foreline trap has
excessive moisture.
Replace filter.
Rotary pump set to gas
ballast.
Switch the gas ballasting off.
Rotary pump requires oil
change.
Change oil.
High mass spectra appears
as a large blotch, or loss of
high mass spectra.
Bad tune.
Run AutoTune.
Drastic change in mass peak
shape for no apparent
reason.
Bad tune.
Run AutoTune.
No spectra, or large blotch.
Bad tune.
Run AutoTune.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......