Clarus SQ 8 MS Hardware Guide
108
Source Maintenance
NOTE: Before beginning this procedure, you may want to have on hand the following source
component kits; Rebuild Kit Part No. N6480080 and N6480081.
CAUTION
ATTENTION
To ensure that the mass spectrometer remains contamination free, wear
lint-free, powder-free nitrile
,
nylon,
or PVC gloves while performing
this procedure.
Pour s'assurer que le spectromètre de masse reste exempt de
contamination, porter des gants de nitrile, de nylon ou de PVC sans
peluches sans poudre lors de cette procédure.
Disassembling
Prepare a clean, uncluttered work area and place a square of aluminum foil with the
shiny side up. Obtain some clean, small containers (for example, small beakers) to
store the small parts as you remove them.
1.
Remove the source by following the
Removing and Returning the Source
on
page 82
2.
Orient the source so that the filament faces you, then locate the filament
retaining clip that holds it in place.
Figure 26 Removing the Filament.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......