Clarus SQ 8 MS Hardware Guide
164
Problem
Probable Cause
Solution
Slowly falling baseline
(from a high initial value).
Split valve left closed during
acquisition.
Open the split valve.
Inadequate purge flow rate.
Increase flow rate.
Poor off for too long.
Reduce purge time.
Low sensitivity.
Dirty source.
Clean the source.
Poor column performance.
Replace column.
Dirty injector.
Replace injector liner.
Source temperature not
optimized.
Set the proper source
temperature.
Detector voltage set too low.
Increase multiplier voltage.
Tune not set correctly.
Run AutoTune.
Poor filament alignment.
Realign or replace filament.
Incorrect column position in
the source.
Reposition the column.