Clarus SQ 8 MS Hardware Guide
80
WARNING
AVERTISSEMENT
Remember, the transfer line was heated and it may take at
least 10 to 20 minutes to cool.
Rappelez-vous, la ligne de transfert a été chauffée et peut
prendre au moins 10 à 20 minutes pour refroidir.
CAUTION
ATTENTION
Avoid venting to air whenever possible. This eliminates the
introduction of oxygen and water vapor into the mass
spectrometer. The Clarus MS should be vented with UHP
nitrogen (99.9995%). Helium should not be used. To properly
connect a source of dry nitrogen to the instrument order the
manifold venting kit (Part No. N6470045).
Évitez d'évacuer l'air autant que possible. Ceci élimine
l'introduction d'oxygène et de vapeur d'eau dans le spectromètre
de masse. Le Clarus MS doit être ventilé avec de l'azote UHP
(99,9995%). L'hélium ne doit pas être utilisé. Pour connecter
correctement une source d'azote sec à l'ordre des instruments, le
kit d'évacuation du collecteur (N ° de pièce N6470045).
Vent the System
1.
Once both the Inlet Line and Source temperatures have dropped below 100 °C,
select
Vent/Vacuum System Off
from the
Options
menu.
2.
The Vent Pump dialog appears. All pumps are turned off.
3.
Click
OK
.
4.
Observe the Vacuum Pressure Gauges status on the Tune window.
5.
The gauge goes to
ZERO
after the turbopump reaches 50 % speed and the vent
valve is opened.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......