Clarus SQ 8 MS Hardware Guide
154
Problem
Probable Cause
Solution
No spectra, or very little
spectra at the low mass end.
Bad tune.
Run AutoTune.
No spectra, not even noise at
a high multiplier voltage.
Loose electrometer cable.
Reset the cable.
Defective electrometer
board.
Call a PerkinElmer service
engineer.
Multiplier near the end of
its lifetime.
Replace the multiplier.
No filament current.
Defective filament.
Replace the filament.
Wavering baseline (by
several hundred counts).
Defective outer source
temperature sensor or a
defective electrometer.
Replace the temperature
sensor.
Call a PerkinElmer service
engineer.
Increasingly setting higher
multiplier voltage settings.
Multiplier near the end of
its lifetime.
Replace the multiplier.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......