Troubleshooting
151
Troubleshooting Chart
Problem
Probable Cause
Solution
Mass Spectrometer will
not turn on (no indication
of power to the
instrument).
AC line cord not plugged into
an AC outlet.
Plug the Mass
Spectrometer
AC line cord
into an AC outlet.
No AC power to the outlet.
Check the outlet.
Fuse blown.
Call a PerkinElmer service
engineer.
Mass Spectrometer is on
but the forepump is not
running.
Forepump is not plugged into
the AC outlet on the rear of
the mass spectrometer.
Plug the forepump line
cord into the mass
spectrometer.
Forepump is not operating
correctly ( mass spectrometer
does not pump down).
Make sure the forepump is
switched on.
Blown fuse in Mass
Spectrometer.
Call a PerkinElmer service
engineer.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......