Maintenance
131
4.
Place the pump on a table. Place a drain container under the drain plug. Raise
the end of the pump opposite the drain plug by putting a block under it.
WARNING
AVERTISSEMENT
If you were running toxic samples, the oil is contaminated as
toxic waste. Handle and dispose of waste oil appropriately.
Si vous couriez des échantillons toxiques, l'huile est contaminée
en tant que déchet toxique. Manipuler et éliminer l'huile usée de
manière appropriée.
5.
Remove the drain plug and allow the oil to drain into the container.
If the pump oil was contaminated, pour clean oil into the filler hole and allow
it to drain until the oil appears clear.
6.
Replace the drain plug, remove the block and reconnect the vacuum system.
7.
Add oil until it reaches the
MAX
mark on the top of the oil level indicator.
Do not overfill.
8.
Replace the oil filler plug by tightening it until it is fingertight.
Do not overtighten the oil filler plug.
9.
After restarting the pump and allowing it to run for a few minutes, recheck the
oil level. If the oil level is below the
MAX
mark, repeat the above procedure by
adding more oil until it reaches the
MAX
mark.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......