Warnings and Safety Information
Temperature, Humidity, and Environment
Operating Conditions
CAUTION
ATTENTION
The Clarus SQ 8 MS is designed for indoor use only.
Le Clarus SQ 8 MS est conçu pour une utilisation à l'intérieur seulement.
CAUTION
Do not operate the mass spectrometer in a Cold Room or a refrigerated
area. Clarus SQ 8 MS operates under the following conditions:
Ambient temperature is 10 °C to 35 °C (50 °F and 95 °F) with a
variability of less than ± 4 °C (± 7 °F).
The Clarus SQ 8 MS will operate safely between 5°C and 40 °C (41 °F
and 104 °F).
If operating at ambient temperatures 10°C and 35 °C, you will need the
water-cooling option for the turbopump.
Ambient relative humidity is 20 % to 80 % non-condensing.
Operating altitude is in the range of -400 to 2000 m (-1,312 to 6,562
feet).
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......