Heated Zones
WARNING
Risk of burns. Never touch a heated Clarus SQ 8 MS transfer
line or a GC injector cap with bare (unprotected) fingers.
AVERTISSEMENT
Risque de brûlures. Ne touchez jamais une ligne de
transfert chauffée Clarus SQ 8 MS ou un capuchon injecteur
GC avec des doigts nuisibles (non protégés).
Heated zones should be treated with caution. For example, the transfer line, injector
caps, and detectors. In addition, the detector cover may get hot, especially if flame
ionization detectors are operated at high temperatures. As a general rule, allow
heated zones to cool before attempting to work in the GC oven, on the transfer line,
on an injector, around the detector areas or inside the Clarus SQ 8 manifold. Proper
cooling of the transfer line may take from 1/2 to 1 hour.
Using Hydrogen, Methane or Isobutane
WARNING
Explosive Hazard. If the hydrogen is turned on without a column
attached to the injector and/or detector fittings inside the oven,
the gas could diffuse into the oven creating the possibility of an
explosion.
If the mass spectrometer is not under vacuum, hydrogen,
methane, or isobutane can fill the vacuum chamber thereby
creating an explosive hazard.
To avoid possible injury, do not turn on the hydrogen unless a
column is attached, all joints have been leak-tested, and the mass
spectrometer is under vacuum with the forepump exhaust
properly vented to a fume hood.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......