Clarus SQ 8 MS Hardware Guide
144
keep the source cleaner, but may increase
fragmentation.
For example, down a little from EI to minimize
fragmentation. 120
°
C is the practical lower limit.
Ion Energy
Approx. 1 or 2
2.
Optimize the amount of reagent gas flowing into the source by using two
heptacosa ions, m/z 452 and 633, which usually produce relative intensities of
65 – 85% and 95% respectively.
Heptacosa can be used to calibrate the m/z range for negative ion CI analyses.
3.
Maximize the peak intensities, then slightly decrease the reagent gas by turning
the knob clockwise 1/8 turn.
Optimize the tuning parameters for maximum intensity.
4.
Save the Tune page parameters by selecting
Save As…
from the
File
menu.
5.
Select
Calibrate Instrument
from the Tune page
Calibration
menu.
6.
Select
heptaneg.ref
from the drop-down menu.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......