Clarus SQ 8 MS Hardware Guide
60
Pumping Down a Turbomolecular Pump Vacuum System
Select
Pump/Vacuum System On
from the
Options
menu on the Tune page.
The menu name will change from
Pump/Vacuum System On
to
Vent/Vacuum System Off
, and the system will begin its pump-down
sequence. Once
OPERATE
is enabled, it remains enabled unless the
Vent/Vacuum System Off
command is given.
Venting the Turbomolecular Pump Vacuum System
1.
Cool the source and inlet to below 100 °C.
2.
Select
Vent/Vacuum System Off
from the
Options
menu on the Tune
page, and confirm that you want to vent the system.
Vacuum Gauge
The single wide range vacuum gauge monitors the system pressure from atmosphere
down to 10
-9
Torr using a combined Pirani/Inverted Magnetron ionization sensor.
Normal operating pressure with 1 mL/min helium for the 255 L/sec turbomolecular
pump is between 9x10
-6
Torr and 2x10
-5
Torr after pump-down and ion source bake-
out. The 75 L/sec turbomolecular pump will operate at somewhat higher pressures,
typically below 4x10
-5
Torr.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......