Clarus SQ 8 MS Hardware Guide
100
Refilling the Reference Gas Vial
It is time to refill the reference gas vial when you lose reference gas peaks intensity
and you do not see a liquid in the reference gas vial, or any time the mass spectrometer
is vented, and the liquid in the vial level appears low.
NOTE: You should check liquid level in the reference gas vial any time you need to remove the
mass spectrometer cover and vent the instrument. Using a tool such as a dental mirror
will help you observe the liquid level in the Reference Gas Vial.
Items Required
•
Lint-free, powder-free nitrile, nylon, or PVC gloves.
•
Pasteur Pipette or 50 µL syringe.
•
Heptacosa (FC43) (Part No. N6212407).
To refill the reference gas vial, follow this procedure:
WARNING
AVERTISSEMENT
Make sure to vent the instrument and turn the power off.
Assurez-vous de ventiler l'instrument et de mettre l'appareil
hors tension.
1.
Ensure that the solenoid has been switched off.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......