Maintenance
91
4.
Using the edge of a wafer scribe, score the outside surface of the column
perpendicular to its length approximately 2 inches from the end and carefully
break it off and discard the cut off piece of column. Jagged or angled cuts should
be avoided. See
Figure 15
for examples of good cuts and bad cuts.
Figure 15
Good cuts and bad cuts
5.
Use a lint-free wipe pre-soaked with a small amount of methanol to wipe the
outside of the column a few times to remove surface contamination.
6.
Place the column onto a clean, lint-free surface on the bench top. Measure exactly
34.4
cm from the outlet end of the column to the front side of the septum.
7.
Confirm that the Source is fully installed in the instrument. If it is not, install
the source into the instrument. See
Removing and Returning the Source
on page 82Error! Bookmark not defined..
8.
Insert the column into the MS Transfer Line tube and carefully slide it
partially
toward the source. Be careful not to move the septum.
9.
With the column
partially
inserted, engage the threaded column nut onto the
Transfer Line tube fitting until it is just finger-tight.
Bad Cuts
Good Cut
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......