Maintenance
141
Figure 45 CI reagent gas adjustment knob.
1.
Carefully turn the delicate CI reagent gas adjustment knob fully clockwise until
you feel it stop.
2.
Select CI Gas On from the Gas menu. A check mark appears next to the option.
NOTE: Always turn on the CI gas before Operate to avoid a pressure surge hitting the
filament.
3.
Click
Press for Operate
and monitor the Penning gauge as you adjust the CI
gas. Observe that mass 16 initially grows larger. As pressure increases in the
ion chamber of CI source, the mass 29 peak will begin to grow. Keep the
pressure below 5 e-4 Torr.
CI Reagent Gas
Adjustment Knob