Troubleshooting
155
Problem
Probable Cause
Solution
Poor or inadequate
sensitivity.
Leaking injector septum.
Replace the septum.
Leak from injector
ferrules.
Tighten/replace ferrules.
Foreign material in the
injector.
Clean the injector.
Peak splitting.
Prevent double injections.
Dry the outside of the injector
needle.
Replace the injector septum.
Injector and column are
more active toward
acid/base compounds.
Install a silanized injector
liner, or silanize the current
injector liner.
Check or replace the injector
packing material, such as
quartz wool.
Cut off the first 30 cm of the
column and rerun the test mix.
If the results do not improve
replace the column.