Troubleshooting
157
Problem
Probable Cause
Solution
Poor sensitivity (correct
amount of sample is reaching
the mass spectrometer).
Column is improperly
positioned in the ion
source.
Reinstall the column and
check the cut at the end
that fits in the source.
Improper tuning, or a
dirty or defective ion
source.
Check the tuning. Increase
the multiplier voltage.
Shut down the system,
remove the inner source,
clean or replace lenses in
the outer source, and
install a new filament.
Loss of resolution (especially
at high mass).
Dirty prequadrupole rods.
Clean the prequadrupole
rods.
Mass assignment drifts.
Large temperature
fluctuations in the
laboratory.
Stabilize the lab
temperature, or isolate the
GC/MS system from large
temperature fluctuations.
Mass assignment incorrect.
Run mass calibration.
Skewed spectra.
Improper scan speed (too
slow for the narrow peaks
produced by capillary
columns).
Increase the scan speed.
Unusually high repeller
voltage.
Repeller dirty.
Clean repeller.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......