Clarus SQ 8 MS Hardware Guide
102
4.
Using a pipette or syringe, add 25 to 50 µL but no more than 50 µL of
Heptacosa (FC43). See the following figure. Fill the bulb.
Never
add more than 50 µL.
Reference
Gas Vial
50 µL max
Gas Tight
Syringe
Figure 23 Filling the reference gas vial.
5.
Re-insert the reference gas vial into the knurled fitting.
Make sure the O-ring is still present and the tapered end of the ferrule faces the
mass spectrometer.
6.
Tighten the knurled fitting with your fingers until fingertight.
7.
Replace the top cover and pump the system down to the proper vacuum.
8.
From the Gas Menu select the Reference Gas Option from the drop down
menu. Leave the Reference Gas Valve open for 60 minutes with Operate off to
pump to pump out the gas from the bulb before tuning the mass spectrometer.
NOTE: On the Tune Page the Gas drop down the select Reference Gas On.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......