Maintenance
95
14.
After initially heating the GC oven through one or two analysis cycles, the
column nut may loosen. Re-tighten the column nut. After re-tightening, the
column nut should not loosen any more.
Optical Column Alignment Using the Optional Plug Handle
and Sight
CAUTION
ATTENTION
To prevent contamination of the mass spectrometer, always wear clean,
lint-free, powder-free nitrile
,
nylon,
or PVC gloves when handling the
part of the capillary column that will be inserted into the MS Transfer
Line. Never touch these parts with ungloved (bare) fingers, as this will
introduce contaminants into the system.
Pour éviter la contamination du spectromètre de masse, porter toujours
des gants de nitrile, de nylon ou de PVC propres, exempts de peluches
et sans poudre lors de la manipulation de la partie de la colonne
capillaire qui sera insérée dans la ligne de transfert MS. Ne touchez
jamais ces pièces avec des doigts non agglomérés, car cela introduira
des contaminants dans le système.
1.
Loosen the large nut on the Transfer Line tube assembly using the 9/16 inch
wrench and pull back the Transfer Line tube approximately 1 to 2 inches.
2.
Remove the source from the Clarus SQ 8 MS instrument. See
Removing and
Returning the Source
on page 82
3.
Insert the optional Plug Handle and Sight (Part No N6480380). Make sure to
line up the red dots on the plug and instrument.
4.
Turn the plug clockwise until the line on the plug and the lock symbol line up.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......