Clarus SQ 8 MS Hardware Guide
118
8.
Tilt the front cover towards you slightly to provide clearance for removing the
lid. It is not necessary to remove the front cover completely from the
instrument.
9.
Use a 5 mm Allen key to loosen the four hex screws that hold the optics
assembly in place. See the following figure.
Figure 32 Removing the screws on the lid and the front cover.
Screw
Screw
Screws
Holding the
Front Cover