Maintenance
111
CAUTION
ATTENTION
Do not allow the methanol and acetone to touch the O-ring on the
source.
Ne laissez pas le méthanol et l'acétone toucher le joint torique sur la
source.
4.
Carefully drain the acetone.
5.
Dry off the source parts using lint-free tissue and/or clean compressed nitrogen
gas to prevent solvents from drying on these parts and leaving a residue.
If you did not use acetone to rinse the parts, wrap them in a clean, lint-free
cloth and bake them in the GC oven at about 80 ºC for about fifteen minutes.
Reassembling
1.
Replace the source parts into the source as shown below.
Remember to insert
the parts into the source only as positioned below.
You may want to insert the trap insulator and trap before you insert the
repeller.
Figure 29 Replacing the source parts.
2.
Align the Lens 3 grabbers with the pin on the source, then turn Lens 3
clockwise to lock it in place. See Figure 27.
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......