Maintenance
97
5.
Re-insert the Transfer Line tube assembly and tighten the large nut.
6.
Slide a 1/16 inch column nut over the outlet end of the column.
7.
With the tapered end facing towards the column nut, slide a graphite/vespel
ferrule over the outlet end of the column. See Figure 14.
8.
Using a wafer scribe, score the outside surface of the column approximately 2
inches from the end and carefully break it off and discard the cut off piece of
column.
9.
Use a lint-free wipe pre-soaked with a small amount of methanol to wipe the
outside of the column a few times to remove surface contamination.
10.
Insert the column into the MS Transfer Line tube and carefully slide it toward
the source.
Figure 20 Inserting the transfer line.
11.
With the column partially inserted, engage the threaded column nut onto the
Transfer Line tube fitting until it is finger-tight.
Column
Summary of Contents for Clarus SQ 8 MS Series
Page 1: ...CLARUS SQ 8 MS Hardware Guide GAS CHROMATOGRAPHY ...
Page 2: ......
Page 3: ...Clarus SQ 8 MS Hardware Guide ...
Page 8: ......
Page 9: ...Warnings and Safety Information 1 ...
Page 10: ......
Page 18: ...Label Location and Content Figure 1 Front View of Clarus SQ 8 T ...
Page 46: ......
Page 47: ...Introduction 2 ...
Page 48: ......
Page 52: ......
Page 53: ...About the Clarus GC MS System 3 ...
Page 54: ......
Page 64: ......
Page 67: ...Maintenance 65 Maintenance 4 ...
Page 68: ......
Page 69: ...Maintenance 67 ...
Page 98: ...Clarus SQ 8 MS Hardware Guide 96 Figure 19 Inserting the Plug Handle and Sight ...
Page 148: ...Troubleshooting 5 ...
Page 149: ...Troubleshooting 147 ...
Page 178: ...TurboMass Software Installation 176 ...
Page 179: ...Index ...
Page 180: ......
Page 183: ......