Sample Programming and Processing
Sample Tubes Validated for CTS
UniCel DxC Systems Instructions For Use A13914
Preparing Samples for Analysis
October 2005
Page 3-5
3
Sample Tubes Validated for CTS
Introduction
Rubber cap tubes can be run directly on the system without removing the cap. The
CTS assembly cuts a small hole in the cap and the sample probe aspirates a sample
directly from the tube. With this procedure, samples are processed faster and safer.
CTS Validated Sample Tubes
Only validated sample tubes can be run on UniCel DxC Systems with CTS. However,
capped and uncapped tubes can be placed on the same reserved rack. Each tube is
scanned to determine if it needs piercing. Refer to the table below for validated tubes.
Reserved Racks for CTS Samples
The Reserved Racks/Obstruct Detect feature allows the user access to the following
features:
Closed Tube Sampling (CTS)
This is an optional feature that allows the system to pierce primary sample tubes. If a
CTS assembly is installed, the operator can enable/disable the CTS assembly. A CTS
Tracking database monitors pierced tubes. If you run a tube again, it will not be
pierced again. The sample ID stays in this database for seven days for a tube that has
been pierced.
Table 3.3
Tubes Validated for Closed Tube Sampling
CTS Configuration
Tubes Validated
for Cap Piercing
Size
1-Blade Thick CTS
Becton Dickinson VACUTAINER
a
with HEMOGARD
a
a
All trademarks are property of their respective owners.
13 × 75mm
13 × 100 mm
16 × 100 mm
Greiner VACUETTE
a
13 × 75mm
13 × 100 mm
Sarstedt S-Monovette
®ab
b
This tube requires a special 5.5 mL rack (P/N A18642).
75 × 15 mm
92 × 15 mm
CAUTION
Before placing validated closed tubes on the UniCel DxC Systems, check the
top of the cap for any residual blood. Residual blood can contaminate the
sample and affect results. If blood is present, remove it by using a cotton-
tipped applicator stick moistened with DI water. When running in the CTS
mode, if tubes off-loaded from the UniCel DxC Systems have water or
droplets of water on the caps, disable the CTS and call Beckman Coulter
Technical Support. Note: Oil on a cap is normal.