Prime
Priming All Subsystems
Utilities
UniCel DxC Systems Instructions For Use A13914
Page 12-4
October 2005
Priming All Subsystems
If the entire group of subsystems, such as all Cartridge Chemistry (CC) subsystems,
needs to be primed, it is easiest to simply select the group of subsystems for prime
instead of having to select each individual subsystem. Follow the steps below to prime
a group of subsystems.
Priming CC Subsystems
The CC Subsystem option allows independent priming of CC Reagent Delivery
Subsystem, CC Sample Delivery Subsystem and Cuvette Wash.
Step
Action
1
From the Utilities dialog box, select <1> Prime.
OR
Type 1 in the Option Number field and press [Enter].
2
Select one or a combination of the following check boxes:
• All CC Subsystems
• All Hydropneumatic Subsystems
• All MC Subsystems
• Prime DxC
• CTS Auto-Gloss (CTS System only)
• CTS Blade Wash (CTS System only).
3
Type the desired number of primes (1–99) in the Number of primes to
repeat field. This does not apply for Hydro primes.
4
Select <Start Prime>. The system status changes to Priming.
5
Upon completion of the requested prime, the number of primes remaining
will be blank and the system status returns to Standby.
CAUTION
If any two of the CC Subsystem items are selected, all three will prime
automatically. Check to make certain that all three areas are ready to be
primed if any two are selected.