UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-31
Common Error Messages and Corrective Actions
0x01FA0008
4
Waste Exit Sump Is
Not Draining At
All.
CAU: Waste Exit Sump
Is Not Draining At All.
Message ID:
0x01FA0008
Description: This error will flag
when the Waste Exit Sump does not
drain within the given drain time
for the sump.
• Make sure Drain Line is not clogged or
pinched.
• Make sure Instrument is not rolled on
top of the Drain Line.
• If the error continues, Contact
Beckman Coulter Technical Support
Center for additional information.
0x01FA0009
4
DI Water Canister
Is Not Filling At
All.
CAU: DI Water
Canister Is Not Filling
At All.
Message ID:
0x01FA0009
Description: Whenever the DI
water canister does not fill within
its given amount of time, it will
time out and flag that it has not
filled.
• Make sure DI water is turned on at the
source and present in the DI Water
Reservoir.
- If Reservoir is full, press the STOP
button and HOME the system.
- If the error continues, contact
Beckman Coulter Technical Support
Center for additional information.
• If Reservoir is empty,
- Make sure the Manual water inlet
valve is turned on.
- Make sure the instrument has not
rolled on top of the water inlet lines.
- HOME and restart the system.
- If the error continues, contact
Beckman Coulter Technical Support
Center for additional information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action