As-Needed/As-Required Maintenance
ISE Flow Cell Cleaning Using Clenz Solution
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-101
10
Step
Action
1
Loosen the two Phillips screws on the ISE module cover and remove cover.
2
Release the locking pin on the ISE module and lift the module until it
clicks.
3
Prepare the flow cell for maintenance by selecting:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <3> ISE Service,
• <Continue>.
4
Remove line #23 from the bottom of the flow cell. (A Luer connector and
small diameter tubing may be found in the Flow Cell Cleaning Kit (P/N
473181)). Connect the open end of the tubing to the bottom of the flow
cell.
5
Slowly inject the Diluted Clenz Solution into the flow cell. Be sure that
liquid completely fills the flow cell and that there are no air bubbles in the
line. DO NOT REMOVE SYRINGE FROM THE LINE.
6
Allow the system to stand idle for 5 minutes. DO NOT REMOVE
SYRINGE FROM THE LINE DURING THIS TIME.
7
Push on the syringe plunger to completely remove the Clenz Solution from
the flow cell.
8
Use back and forth plunger motion to scrub the flow cell with air.
9
Disconnect the tubing from the bottom of the flow cell.
10
Reconnect line #23 to the bottom of the flow cell.
11
Once the necessary cleaning has been completed, select <OK> to close the
dialog box.
NOTICE
Reagents will immediately prime through the flow cell.
While the flow cell is priming, verify there are no fluid leaks from the
chloride electrode port.
12
After the pop-up message appears, select <Close> to exit the ISE Service
screen.
13
Rinse the syringe, connector, and tubing with deionized water. Dry and
store for future use.
(1 of 2)