![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 477](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768477.webp)
UniCel DxC Systems Instr
u
ctions F
o
r Use A
13914
T
roub
leshooting Calibr
ation and Result Errors
October 2005
13-57
Common Error Messages and Corrective Actions
0x05DC01F8
4
CC Reagent Mixer
Motion Error
CAU: Homing Error for
CC Reagent Mixer.
Message ID:
0x05DC01F8,
Subsystem: boot,
homing command
Description: This error occurs
when there is a motion error with
the Reagent Mixer, either Rotary or
Vertical. The parameter in the event
log entry will show one of the
following:
• "AttachRgtMixer" – error
occurred attaching to smart
module 0xC1.
• "RecoverRgtMixer",
"EnergizeRgtMixerVer",
"RaiseRgtMixer" – error
occurred during a vertical
movement toward the home
sensor flag.
• "HomeRgtMixerVer" – error
occurred during a vertical
movement away from the home
sensor flag.
• "EnergizeRgtMixerCrn",
"MoveRgtMixerToSensor" –
error occurred during a horizontal
movement toward the hardware
sensor.
• "HomeRgtMixerCrane" – error
occurred during a horizontal
movement.
• Raise canopy and check for obvious
obstruction of the CC reagent mixer.
• Press the STOP button and HOME the
system. While system is homing,
observe motion of the CC reagent
mixer. The mixer should complete its
operations in a smooth manner.
• If the error continues, check the event
log to determine the specific subsystem
error.
• Perform the appropriate vertical or
horizontal CC reagent mixer
alignment.
• Contact Beckman Coulter Technical
Support Center for additional
information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action