As-Needed/As-Required Maintenance
Replacing the Sodium Measure/Reference Electrode
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-76
October 2005
8
Loosen the flow cell retaining captive screw (1) located the top of the flow
cell. Pull flow cell assembly forward.
Figure 10.50
9
Unscrew (counterclockwise) the electrode retainer nut (1) and withdraw
the electrode (2) from the flow cell.
Verify that the quad-ring from the removed electrode has not been left in
port opening.
Figure 10.51
10
Remove the retainer nut from the old electrode and install on new
electrode.
Step
Action
, continued
(3 of 4)