As-Needed/As-Required Maintenance
Washing CC Reagent Cuvettes with CCWA
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-116
October 2005
Washing CC Reagent Cuvettes with CCWA
CC Reagent Wash All Cuvettes is an automated procedure that cleans all the cuvettes
on the Reaction Carousel using a cartridge filled with cleaning solution. This
procedure should be performed on dirty cuvettes that require additional washing
beyond the Washing All Cuvettes procedure. This routine will take approximately 38
minutes.
Material Required:
• CCWA cartridge (PN 657133)
NOTICE
System will not autoprime in
Maintenance
mode. After completing
maintenance procedure, or when a procedure is aborted, exit Maintenance
menu to resume autoprime. Return the system to
Standby
.
Step
Action
1
Load the CCWA cartridge onto the system by selecting:
• the Rgts/Cal icon from the menu bar,
• An empty chem position to load,
• Load [F1].
2
This procedure is accessed by selecting:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <9> CC Reagent Wash All Cuvettes (or Type 9 in the Procedure No.
field and press [Enter]).
Follow all instructions listed on the screen.
3
Select <Continue> to perform the procedure.
If desired, exit this procedure at any time before it is completed by
selecting:
• <Stop> on the instrument; wait for system to finish processing,
• <OK> to clear popup message,
• <Cancel> to Exit the procedure screen,
• the Instr Cmd icon from the menu bar,
• <1> Home.
4
Select <Close> at the completion of this procedure to exit the pop-up
Result window.
(1 of 2)