Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-42
October 2005
0x02620033
4
10 Cuvettes Have
failed Water Blank
CAU: 10 Cuvettes Have
Failed Water Blank
Message ID:
0x02620033
Subsystem: RaSched
Description: This error can be
flagged for the following:
• A cuvette's wavelength
absorbance is over 350mA
(Absorbance) for all wavelengths
between 340 to 700 nanometers.
• A cuvette's wavelength
absorbance is over 200mA
(Absorbance) for the 940
nanometer wavelength (Pro
System only).
• PSM Error.
• LPIA Failure.
• Remove wash station head and inspect
probes for blockage and wiper for
wear.
• Check wash station wiper alignment.
• Check wash station alignment.
• Remove reaction carousel and check
cuvette cleanliness.
Non-resolution of problem may indicate
photometer issues. Contact Beckman
Coulter Technical Support Center for
additional information.
0x02620034
4
CC Reagent probe
disabled due to
level sense errors
CAU: CC Reagent
Probe Disabled Due To
level Sense Errors
Message ID:
0x02620034
Subsystem: RASched
Probe: probe
Description: Multiple reagent
cartridges have level sense errors.
• HOME the system.
• Check reagent cartridges for bubbles.
• Perform Reagent Probe to Cartridge &
Reagent Probe Height in Cartridge
Alignments.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action