Temperatures
Introduction
UniCel DxC Systems Instructions For Use A13914
System Status and Commands
October 2005
Page 11-5
11
Temperatures
Introduction
The actual temperature of each of the various components is shown along with:
• the valid range.
• an indication of when a parameter is out of limits.
A yellow highlight indicates a warning condition and a red highlight indicates an error
condition.
Displaying Temperature Status
Follow the steps below to monitor instrument components (e.g., reaction carousel,
cups, cuvette wash).
NOTICE
While the system is in the Stop state, temperatures will not be accurate. If the
system boots directly into the
Stopped
state, no temperatures will be shown for
the systems that caused the Stop state. If the system goes to
Stopped
from
Standby
or
Running
, the temperatures shown will reflect the previous state and
will NOT be updated while the system is stopped.
Step
Action
1
Select the Status icon from the menu bar.
2
Select Temp [F3].
3
A summary of the monitored areas appears.