Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-44
October 2005
0x0262003F
4
Too many tubes not
venting correctly,
sampling for the
CC probe will be
stopped
CAU: CC Sample Probe
disabled due to Excess
Vacuum Errors.
Message ID:
0x0262003F
Subsystem: ccScheduler
Description: Multiple tubes had
excessive vacuum in thick stopper
closed tubes. CC Sample Probe
Disabled.
• Check cap piercer for broken blade.
• Perform CC ODC Alignment.
0x02620040
4
Too many tubes not
venting correctly,
sampling for the
CC probe will be
stopped
CAU: CC Sample Probe
Disabled due to Excess
Pressure Errors.
Message ID:
0x02620040
Subsystem: ccScheduler
Description: Multiple tubes had
excessive pressure in thick stopper
closed tubes. CC Sample Probe
Disabled.
• Check cap piercer for broken blade.
• Perform CC ODC Alignment.
0x02660001
4
No Diluent
Cartridge
CAU: No diluent
cartridge
Message ID:
0x02660001
Subsystem: RASched
Chem Name: chem id
Description: Diluent cartridge
must be loaded to performed
requested tests. Tests requiring the
diluent cartridge will be
incomplete.
• Load a DIL1 Cartridge.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action