Common Error Messages and Corrective Actions
0x13A00057
10
Required levels for
calibrator not
loaded. Rack will
be unloaded from
the sample
carousel.
Rack: rack ID
Calibrator:
calibrator name
WLM: Subsequent
calibrator levels not
loaded for multi-rack
calibrator on rack rack
ID (File Name:Line
Number)
Description: Not all calibrator
levels were loaded or were loaded
out of order.
• Reload calibrator racks.
• Make sure bar codes are readable.
• If there are 4 calibrators or less, they
must all be in the same rack.
• If more than 4 levels are required,
levels 1–4 must be in the first rack
loaded and the next rack loaded must
contain the remaining levels.
0x13A00059
10
Scanned barcodes
in rack #rack ID
contain invalid
characters. Rack
will unloaded from
the sample
carousel.
WLM: Barcodes contain
invalid chars.
Rack: rack ID
[1:sample ID]
[2:sample ID]
[3:sample ID]
[4:sample ID]
Description: Sample bar code
contains illegal characters.
• Check the manual for illegal
characters.
• Make sure the complete bar code is
visible and there are no smudges or
wrinkles.
0x13A0007E
10
Module: xxxxxx is
disabled due to a
system error. All
module functioning
has now ceased.
When the problem
is fixed, enable the
module through
Instrument
Commands.
WLM: Received request
to disable module:
Module (File Name:Line
Number)
Description: Specified module has
been disabled automatically. This
will normally have severity of IM.
• Enable the module under Instrument
Commands, Enable/Disable. If the
problem continues, it could be a valve
or vacuum problem.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action