Error Code – Definitions
Result Error
Troubleshooting Calibration and Result Errors
UniCel DxC Systems Instructions For Use A13914
Page 13-28
October 2005
Result Error
This error occurs when a result does not fit into a specific calibration curve. Rerun the
sample for verification. This error may indicate sample interference issues. It may
also indicate issues with photometer performance. Contact Beckman Coulter
Technical Support for additional information.
Sample in Wrong Rack
The requested tests have been programmed in the wrong rack when pre-programmed
reserved racks have been specified for requested tests. Refer to
in CHAPTER 4, System Setup Options in this manual for additional
information on reserved racks.
Sample Signal Noise
The difference between the high and low values of the eight sample readings has
exceeded the limits allowed. Units are measured in ADCs and are a measurement of
system noise observed during the sample cycle. Consistent flags may indicate poor
system grounding.