![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 440](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768440.webp)
Result Errors and Codes
Troubleshooting Calibration and Result Errors
UniCel DxC Systems Instructions For Use A13914
Page 13-20
October 2005
Results Suppressed
INI RATE LO
IL
CC chemistries, initial rate too low.
MC chemistries, a stirrer failure.
Results Suppressed
INIT ADC HI
HI
Initial ADC error high.
Results Suppressed
INIT ADC LO
LI
Initial ADC error low.
Results Suppressed
INIT COND HI
CH
Initial conductance high.
Results Suppressed
INIT COND LO
CL
Initial conductance low.
Results Suppressed
K REQ
IK
Bad K value.
Results Suppressed
NA/K REQ
IN
Bad NA value.
Results Suppressed
OIR HI
DH
Out of instrument range high.
Results Suppressed
OIR LO
DL
Out of instrument range low.
Results Suppressed
ORDAC HI
OH
Out of instrument range ORDAC high.
Results Suppressed
ORDAC LO
OL
Out of instrument range ORDAC low.
Results Suppressed
URDAC HI
GH
Under range URDAC high.
Results Suppressed
URDAC LO
GL
Under range URDAC low.
Results Suppressed
ORR HI
UH
Out of reportable range high.
Results Suppressed
ORR LO
UL
Out of reportable range low.
Results Suppressed
ORR O HI
UO
Out of ORDAC reportable range high.
Results Suppressed
OVERFLOW
OF
Number overflow error.
Results Suppressed
REF NOISE
DR
Reference signal noise (ISE only).
Results Suppressed
RX ABS HI
HR
Reaction absorbance high.
Results Suppressed
RX ABS LO
LR
Reaction absorbance low.
Results Suppressed
RX ERR
RE
Reaction error.
Results Suppressed
RX MAX DEV
RO
Reaction outlier (maximum deviation).
Results Suppressed
RX MEAN DEV
RN
Reaction noise (mean deviation).
Results Suppressed
RX NOISE
EA
A
Erratic ADC (Cup Chems only).
Results Suppressed
RXN RATE HI
RH
Reaction rate high.
Results Suppressed
RXN RATE LO
RL
Reaction rate low.
Results Suppressed
SAMP NOISE
DS
Sample signal noise (ISE only).
Results Suppressed
SAMP REF DRIFT
ES
Excessive reference drift – sample to
sample.
Results Suppressed
SUB DEPL
SD
Substrate depleted.
Results Suppressed
TEMP ERR
TM
T
Temperature error (MC only).
TEST STOPPED
Test was aborted, system issue.
TOO MANY TESTS
Too many tests on CC.
Table 13.9
Result Errors and Codes
, continued
Result Displayed on Report
Printed in
Remarks Area
Host
Code
Inst.
Code
Description
(3 of 3)