Three-Month Maintenance
Replace Syringe Plungers
UniCel DxC Systems Instructions For Use A13914
Maintenance
October 2005
Page 10-45
10
Figure 10.27 MC and CC Sample Syringes (100 µL)
Preparing for Maintenance Mode and Removing the Plunger
Before removing the plunger, the plunger rods must be fully extended to the bottom of
the syringe barrels. Follow the steps below to activate the automated Syringe Rod
Replacement procedure.
1. MC Sample Syringe
2. CC Sample Syringe
NOTICE
The system status must be in
Standby
to access the
Maintenance
mode.
Step
Action
1
Select:
• the Utils icon from the menu bar,
• <2> Maintenance,
• <7> Syringe Rod Replacement.
Select <Continue> to prepare the syringe rod assemblies for maintenance.
OR
Select <Cancel> to immediately return to the Maintenance menu without
performing syringe rod maintenance.
(1 of 3)
A015943P.EPS
1