Pause/Resume Waste B
Introduction
UniCel DxC Systems Instructions For Use A13914
System Status and Commands
October 2005
Page 11-21
11
Pause/Resume Waste B
Introduction
The Waste B option allows for some of the waste to be collected in an external bottle
instead of going immediately down the drain. It is only necessary to use the Pause
Waste B option if your system is configured with a Waste B collection bottle. Pause
Waste B prevents the instrument from filling the collection bottle to allow for
exchange of bottles. Resume Waste B is used to return to normal operation after
Waste B removal has been paused or to recover from certain flood conditions.
How to Pause Waste B
Waste B removal will be paused until Resume Waste B is selected. The Instrument
Commands screen remains so that Resume may be selected. Pause Waste B will be
gray, to indicate it is already paused.
How to Resume Waste B
The system will return to its normal operational state. The Instrument Commands
screen will close automatically.
NOTICE
If the Waste B option is paused and the Waste B exit sump is full, the system
will not run until the Resume Waste B option is selected.
Step
Action
1
Select the Instr Cmd icon from the menu bar.
2
Select <5> Pause Waste B.
Step
Action
1
Select the Instr Cmd icon from the menu bar.
2
Select <6> Resume Waste B.