CTS Wick Replacement Procedure
Install the Wick Clip Assembly and Blade Assembly
Maintenance
UniCel DxC Systems Instructions For Use A13914
Page 10-130
October 2005
11
For a DxC 600/800 system, place the rack into the autoloader with no
sample programming. Press the RUN button to pierce the tube caps.
Inspect the bottom of each cap to confirm that the cap was pierced.
12
For a DxC 600i system, loosen the two Phillips screws on the cover of the
DxC Load Area and remove the cover. Place the rack into the DxC Load
Area with no sample programming.
Note: Do not load the rack into the DxC 600i Load Area.
Loosen the screw and open the cover of the DxC Load Area RUN button
(located to the right of the DxC 600i Load Area). Press the DxC Load
Area RUN button to pierce the tube caps. Inspect the bottom of each cap
to confirm that the cap was pierced. Close the DxC Load Area RUN
button cover and tighten the screw. Replace the DxC Load Area cover.
Step
Action
, continued
(2 of 2)