leshooting Calibr
ation and Result Errors
October 2005
13-51
Common Error Messages and Corrective Actions
0x04B20026
4
Sample tube did not
vent correctly.
Remove cap and
reload tube.
Sample ID: sample
id
Rack: sample rack
id
Position: sample
position, MC
Sample Probe
CAU: MC Sample Tube
Excess Vacuum
Message ID:
0x04B20026
Sample ID: sample id
Rack ID: sample rack id
Pos: sample position.
Description: Probe detected
excessive vacuum in thick stopper
closed tube on second attempt.
Sample is aborted.
• Uncap tube and rerun.
• If occurs consistently for many
samples, check cap piercer for broken
blade.
• Perform MC ODC Alignment.
0x04B20027
4
Sample tube did not
vent correctly.
Remove cap and
reload tube.
Sample ID: sample
id,
Rack: sample rack
id,
Position: sample
position, MC
Sample Probe
CAU: MC Sample Tube
Excess Pressure
Message ID:
0x04B20027
Sample ID: sample id,
Rack ID: sample rack
id,
Pos: position.
Description: Probe detected
excessive pressure in thick stopper
closed tube on second attempt.
Sample is aborted.
• Uncap tube and rerun.
• If occurs consistently for many
samples, check cap piercer for broken
blade.
• Perform MC ODC Alignment.
0x04B3000F
4
ISE DAC setting
Failed
Chem Name: chem
id
CAU: ISE DAC Setting
Failed
Message ID:
0x04B3000F
Chem Name: chem id
Description: ISE DAC Setting
Failed for the specified chemistry.
• Make sure the ISE module is in the
down position, the cover is in place the
reagent door is closed.
• Check for reagent volume in bottles or
flow cell leaks.
• Check electrode connections.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action