Calibration Override
Introduction
Reagent Load/Calibration
UniCel DxC Systems Instructions For Use A13914
Page 5-26
October 2005
Calibration Override
Introduction
The Calibration Override option allows the system to override a failed calibration and
obtain results based on the failed calibration factors. Calibration Override is available
only when an attempt to calibrate the chemistries has been previously made and failed,
the status Cal Failed is shown. A chemistry with any other status, including Cal
Required, cannot be overridden. The calibration can be overridden while the system
status is Standby, Disabled or Running.
Overriding a Calibration
Follow the steps below to override a calibration.
Removing Cal Overridden Status
The Cal Overridden status can only be removed by successfully recalibrating the
chemistry or by selecting Chem Bypass. A flag indicating that a particular chemistry
has been overridden will appear in the Instrument Code area on the appropriate
Summary or Patient Chartable Results report. Although a Cal Overridden status can
be bypassed, the Cal Overridden flag is only removed by a successful calibration.
NOTICE
Overriding a calibration is not usually justified. However, there may be an
emergency situation when a delay caused by calibration is unacceptable to the
laboratory. The magnitude of error which is deemed acceptable when
overriding a failed calibration is totally under the control of the laboratory.
Therefore, the Override function should be used with caution.
Step
Action
1
Select the Rgts/Cal icon from the menu bar.
2
Select the Chem position to override.
3
Select Options [F6].
4
From the Cal Options dialog box, select <7> Calibration override.
OR
Type 7 in the Option Number field and press [Enter].