Common Error Messages and Corrective Actions
0x02620024
2
CC Sample Mixer
Motion Error
CAU: CC Sample Mixer
Motion Error
Message ID:
0x02620024
Subsystem: RASched
Action:
Description: This error will occur
when there is a motion error within
the Sample mixer crane. These
errors will occur while in
"Standby," "Priming," or during
runtime. This error will not pop up
during a home. This can be a rotary
or vertical motion error.
• Make sure mixer has the necessary
room to move.
• Make sure that the wiring and cabling
are secured properly.
• Make sure mixer paddle is seated
firmly in it's housing.
• HOME and restart the system.
• If mixer appears out of position,
perform rotary or vertical alignment.
• If motion error continues, contact
Beckman Coulter Technical Support
Center for additional information.
0x02620025
2
CC Sample Syringe
Motion Error
CAU: CC Sample
Syringe Motion Error
Message ID:
0x02620025
Subsystem: RASched
Action Name: action
name
Description: This error occurs
when there is an error within the
Sample Syringe Subsystem. This
includes both the Shear valve Motor
and the Lead Screw Motor.
• Make sure syringe and plunger are
properly seated at top and base of
syringe.
• Check for obstruction of syringe
movement.
• Perform CC Syringe Home Alignment.
• If the error continues, contact Beckman
Coulter Technical Support Center for
additional information.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action