Error Code – Definitions
Out of Instrument Range High/Low
Troubleshooting Calibration and Result Errors
UniCel DxC Systems Instructions For Use A13914
Page 13-26
October 2005
Out of Instrument Range High/Low
The recovered value exceeds the value that the instrument will report.
Out of ORDAC Range High/Low
A reaction exceeded the range of ORDAC (Over Range Detection and Correction)
values that the instrument will report.
If a sample has an ORDAC LO message and Manual ORDAC is on, rerun the sample
with manual ORDAC off.
If a sample has an ORDAC LO message and Auto ORDAC is on:
1. Dilute the sample 1:2.
2. Program the sample as a 1:2 dilution.
3. Rerun the sample.
If the system gives an answer, report this result. If the result again gives an ORDAC
LO, do a larger dilution and rerun for the applicable dilution.
Out of ORDAC Reportable Range
Result exceeds the operator defined ORDAC range high. Refer to
in CHAPTER 4, System Setup Options in this manual for additional information.
Out of Reportable Range High/Low
Result exceeds the operator defined reportable range high or low. Refer to
in CHAPTER 4, System Setup Options in this manual for additional
information.
Probe Obstruction Detected
The system detected a block in the sampling system when sampling the requested
samples. Check affected samples for clots, then reload. If issue still persists, refer to
the UniCel DxC Synchron Clinical Systems Reference Manual, CHAPTER 5,
Advanced Troubleshooting, CC Hardware Symptoms or MC Hardware Symptoms for
additional investigation.
Reaction Absorbance High/Low
The mean absorbance measured during the reaction spin cycles exceeded limits. The
units are measured in absorbance and will usually characterize sample integrity. If a
high flag is triggered, dilute the sample and rerun. If a low flag is triggered, rerun
sample for verification, except for TDMs. TDMs with low reaction absorbance flags
should be diluted and rerun.