![Beckman Coulter UniCel DxC Synchron 600 Instructions For Use Manual Download Page 484](http://html.mh-extra.com/html/beckman-coulter/unicel-dxc-synchron-600/unicel-dxc-synchron-600_instructions-for-use-manual_2719768484.webp)
Common Error Messages and Corrective Actions
T
roub
leshooting Calibr
a
tion and Result Errors
UniCe
l DxC Systems Instr
u
ctions F
o
r Use A13914
13-64
October 2005
0x139F0068
7
CTS Tracking
Information can
NOT be shared
between all
instruments. When
you move a tube to
a different
instrument, remove
it's cap. You will be
notified when full
CTS tracking is
restored.
STM: TS-CTS Tracking
System disconnect
detected.
Message ID:
0x139F0068
Description: This error will occur
whenever 1 or more instruments is
no longer communicating Cut Tube
Tracking information with the other
networked instruments for any
reason.
NOTICE
TS-CTS stands for
Thick-Stopper, Closed
Tube Sampling
• If a networked system has been
shutdown and rebooted, ignore this
message. If not, check power to all
Microhubs and network cable
connections.
• Select "Status", then [F8], CTS Status
to determine which instrument is no
longer connected.
CAUTION
To check a pre-cut tube is not
cut a second time, any pre-cut
tube moved to the
disconnected system should
have its cap removed prior to
loading.
• Once the connections have been
re-established, all Cut Tube Tracking
databases will re-synchronize and it
will no longer be necessary to remove
caps.
Table 13.10
Error Messages and Corrective Actions
, continued
ID
Class #
Pop-up
Event Log
Description
Corrective Action