Pause
When NOT to Use Pause
System Status and Commands
UniCel DxC Systems Instructions For Use A13914
Page 11-14
October 2005
When NOT to Use Pause
Do not use Pause when there is danger involved or when the situation can be handled
with another function.
How to Pause
Note: Press RUN to continue the run after a Pause.
Table 11.3
Alternatives Instead of Pause for Some Situations
Situation...
Instead of Pause...
Imminent danger to a person or
the instrument.
press the STOP button or turn off power.
Reagent needs to be loaded
while the system is running.
perform a reagent load as usual and the system will
automatically interrupt reagent additions to allow
the reagent load to continue. Upon completion of
the reagent load, the system will resume running
samples.
Sample on the sample carousel
is needed immediately.
from the Main screen select the rack with the
sample needed and select Unload [F2]. Note that
any tests that have been started for the samples in
the rack that have not yet had sample added will be
aborted. The rack, however, will be available
within approximately 40 seconds.
Step
Action
1
Select the Instr Cmd icon from the menu bar.
2
Select <2> Pause.
3
From the Pause dialog box, select <1> MC Only, <2> CC Only or <3>
Both MC and CC to pause the related part of the instrument.
OR
Type 1, 2 or 3 in the Option Number field and press [Enter].
NOTICE
Do not request a
Pause
and leave the system sitting for a long period of time
without washing the cuvettes. If cuvettes are left sitting with reactant,
evaporation will occur, leaving encrustation on the sides of the cuvette that
may require washing by hand.